Performance measurements of the electromagnetic calorimeter and readout electronics system for the DarkQuest experiment
Journal article   Peer reviewed

Performance measurements of the electromagnetic calorimeter and readout electronics system for the DarkQuest experiment

Aram Apyan, Christopher Cosby, Yongbin Feng, Alp Gelgen, Stefania Gori, Philip Harris, Xinlong Liu, Mia Liu, Petar Maksimovic and Cristina Mantilla-Suarez
Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, Vol.1080, p.170792
11/01/2025

Abstract

Calorimeter Dark sectors Test beam
This paper presents performance measurements of a new readout electronics system based on silicon photomultipliers for the PHENIX electromagnetic calorimeter. Installation of the lead-scintillator Shashlik style calorimeter into the SeaQuest/SpinQuest spectrometer has been proposed to broaden the experiment’s dark sector search program, an upgrade known as DarkQuest. The calorimeter and electronics system were subjected to testing and calibration at the Fermilab Test Beam Facility. Detailed studies of the energy response and resolution, as well as particle identification capabilities, were performed. The background rate in the actual experimental environment was also examined. The system is found to be well-suited for a dark sector search program on the Fermilab 120 GeV proton beamline.